| Instrument Calibrated/ Measurement
Parameter |
Range |
Best measurement capabilities expressed
as an uncertainty (±) |
Gas measurement
(Site Calibration)
Ozone concentration |
0 to 200 pphm |
3% of measured value |
Mass measurement
(Site-Calibration)
Analytical balance
Weighing device |
0 to 200 g
200 g to 4 000 g
4 000 g to 6 000 g
6 000 g to 34 000 g |
0.4 mg
3 mg
30 mg
450 mg |
Force measurement
(Site calibration)
Tension
Compression |
1 N to 500 N
0 to 5 kN
5 kN to 10 kN
10 kN to 100 kN
100 kN to 300 kN
300 kN to 5 000 kN |
0.01 % of measured load
0.14 % of measured load
0.28 % of measured load
0.30 % of measured load
0.30 % of measured load
0.74 % of measured load |
| Instrument Calibrated/ Measurement
Parameter |
Range |
Best measurement capabilities expressed
as an uncertainty (±) |
Hardness measurement
IRHD Normal (Dead load)
IRHD (Spring load)
JIS A "
Shore A " |
0 to 22 N
0 to 22 N
0 to 22 N
0 to 22 N |
0.20 % of measured load
0.44 % of measured load
0.81 % of measured load
0.70 % of measured load |
Profile Projector
Indentor geometry
- Diameter of indentor
- Indentor extension
Die cutter measurement
- Width
- Length of parallel portion
|
0 - 90 mm
0 - 90 mm |
0.004 mm
0.004 mm |
Temperature Heat Enclosure
Type K Thermocouple with Recorder
|
40°C - 200°C
40°C - 200°C
|
0.6°C
0.4°C
|
Plastimeter
Load
Displacement Test
Timing |
100N
0.2mm - 1.0 mm
15.0 sec |
0.1 N
0.06 mm
0.08 sec |
Mooney Viscometer
Torque
Platen Temperature
Rotor Speed |
0 to 100 N
(0 to 100 Mooney Unit)
100°C
2 rpm
|
0.4 N
0.2°C
0.1 rpm
|
Rheometer (Monsanto R100)
Torque
Timing
Platen Temperature |
42 Ib.in
0 min to 40 min
(0 sec to 2400 sec)
160°C
177°C |
0.3 Ib.in
0.21sec
0.2°C
0.2°C |