Product Testing & Calibration Unit  
        About Us | Rubber Products | Bearings | Seals | Condoms | Exam Gloves | Surgical Gloves |
Household & Industrial Gloves | Tyre Testing | Calibration Services/Sales | Others | Contact
  Product Testing &
  Calibration Unit

Materials Characterisation
  Unit
    Product Testing & Calibration Unit

RRIM Relaxed Modulus Tester Model M403
RRIM Relaxed Modulus Tester Model M403

The RRIM Relaxed Modulus Tester Model M403 is a rapid and accurate instrument for measuring the modulus of rubber at a fixed elongation. This property is a very useful indicator of the state of vulcanisation in rubbers.

The tester is simple in its construction and operation. The load, in Newtons, is measured by a load cell and shown on a indicator. This load is held after 1 minute and the operator is prompted by a LED indicator light to take the reading on the DPM, after this time has elapsed, to obtain the relaxed modulus after 1 minute.

Specification
Load Cell Rated Capacity 44.4 N
Load Ranges 0 - 40 N
Load Weighing Accuracy ± 0.01 N
Power Requirement AC 210/240V, 50 HZ Single Phase
 
Dimensions
Width 200 mm
Depth 150 mm
Height 560 mm
Weight 5.14 kg

Instrument Description
The RRIM Relaxed Modulus Tester Model M403 incorporates a highly sensitive load cell for detecting the load on the specimen under test. This load cell is attached at the top of a rigid single column load frame which is fixed at its base to a casing which contains the electronics. The load indicator, balance and calibration keys for the load cell weighing system, 1 - minute LED indicator, switch for load range selection and reset button are on the front of this casing.

The tester handles both square tab-ends dumb-bell and rings test pieces. The primary spring loaded grips are for use with square tab-ends dumb-bell test pieces. For use with ring test pieces, hook grips can be attached to this basic facility.

The maximum separation of the primary grips is set at 200 ±0.5 mm to give an elongation of 100% on standard 100 mm long square tab-ends test pieces. The length of the hook grips can be adjusted in accordance with the diameter of the ring test piece to give 100% elongation.

The load weighing system includes a solid state load cell signal conditioning amplifier, a full scale load ranges of 40N. The amplifier controls provide for balancing and calibration of the load weighing system to give accurate load readings.

Applications
  • Vulcanised rubber
  •   Relaxed Modulus at 100% elongation
  • Latex films/gloves
  •   Prevulcanised relaxed modulus, PRM
    Back

    The Government and Malaysian Rubber Board shall not be liable for any loss or damage caused by the usage of any information obtained from this site.