Measurements of Total Extractable Proteins in Latex Gloves: A Comparative Study of the RRIM and ASTM Tests

Introduction

In view of the latex protein allergy problem relating to NR latex products, especially gloves1 , there is a need to evaluate the allergic potential of these products. However, till to-date, a universally standardized test for doing so is still lacking. Many different methods2-9 are being used in various laboratories in different parts of the world. The methods most commonly adopted in manufacturing countries, particularly Malaysia, are the RRIM test (which has recently become a Malaysian Standard Test)7 and the ASTM test8 , both for the determination of total extractable protein content of latex products.

While the RRIM test has been employed for as long as the beginning of the allergy problem, the ASTM was established a few years later. It is often asked how different are the two sets of values so obtained and if they are correlated. The present study has therefore been undertaken to compare these values and to see if any relationship exists between them. Furthermore, in view of the ageing requirement by the FDA for "Low Protein Labelling Claim" for latex gloves10, this effect on total extractable proteins is also investigated.

Materials and Methods

Preparation of gloves pieces.
Square pieces, of dimension 7 cm x 7 cm, were cut from palm area of each glove sample. Care was taken to avoid any contamination.

Accelerated ageing
Glove pieces were subjected to accelerated ageing at 70°C for 7 days, according to ASTM 3578-91.

RRIM modified Lowry test7 (Malaysian Standard Protein Test MS 1392-96P)
Cut pieces of each test sample were extracted in 0.01M phosphate buffered saline (pH 7.4) at 23°C for 3 hours. Extract was clarified by centrifugation, to sediment any particulate matter, such as powder. Precipitation of proteins was carried out using trichloroacetic and phosphotungstic acids. This was followed by further centrifugation at 10,000 x g for 30 minutes. The resulting pellet was redissolved in 0.2 M sodium hydroxide, and their concentration determined by the Lowry microassay. Absorbance readings, recorded at 750 nm, were calibrated against standard bovine serum albumin (BSA). Results were expressed as EPRRIM in µg/g of gloves.

ASTM Protein Test8.
Proteins were extracted in distilled water at 37°C for 2 hours. Extract was centrifuged and the proteins were precipitated using deoxycholate, trichloroacetic and phosphotungstic acids. Sedimentation of the protein pellet was carried out by centrifugation at 6000 x g for 15 minutes. The precipitated proteins were redissolved in 0.01 M sodium hydroxide, and their concentration determined by Lowry microassay according to procedure set out in the BioRad DC protein kit. Absorbance was recorded at 750 nm and readings calibrated against standard ovalbumin. Results were presented as EPASTM, in µg/g of gloves.

Results

Total Extractable Protein Contents

A total of 90 commercial gloves were extracted and tested. Results are summarized in Table 1. It can be seen that total extractable protein content, EPRRIM, ranged from as high as 1326 µg/g to as low as to 19 µg/g. Their median and overall mean values were 505 µg/g and 519 µg/g respectively. Similarly, values of EPASTM, varied from 1377 µg/g to 36 µg/g, with a median of 372 µg/g and a mean of 423 µg/g. Comparison of the overall means indicated EPASTM was 18.5% lower than that of the EPRRIM, while in the case of the medians, the decrease was 26.3%.

Closer examination of the data at different EPRRIM ranges, revealed that means of EPASTM read 13.8% - 25% lower than those of EPRRIM when the latter values were greater than 100 µg/g. However, at EPRRIM of 100 µg/g and lower, the mean of EPASTM indicated 17.6% higher values than that of EPRRIM (Table 2.)

Comparison of data based on different EPASTM ranges, on the other hand, was found to vary somewhat as shown in Table 3. Although lower EPASTM values than those of EPRRIM was still apparent, the difference was more marked at the low range. Nevertheless, statistical analysis of the two sets of data showed that they are very closely related, with the coefficient of correlation, r = 0.93, P < 0.001 (Figure 1).

TABLE 1. COMPARISON OF EXTRACTABLE PROTEIN CONTENTS BY RRIM AND ASTM METHODS FOR 90 GLOVE SAMPLES.

Item EPRRIM
µg/g)
EPASTM
µg/g)
%
diff.
Range 19 to 1326 36 to 377 -
Median 505 372 26.3
Mean 519 423 18.5
S.d. 407 348 -

S.d. - Standard deviation

TABLE 2. MEANS OF EPRRIM AND EPASTM AT DIFFERENT EPRRIM RANGES FOR 90 GLOVE SAMPLES.

EP
(µg/g)
EPRRIM range (µg/g)
100 & less >100 to 400 >400 to 700 >700 to 1000 >1000
No. of samples 20 21 16 18 15
Corresponding
EPASTM
Range


36 - 107


49 - 408


219 - 673


360 - 1377


557 - 1238
EPRRIM
Mean(S.d.)
51
(28)
212
(63)
551
(86)
846
(83)
1148
(102)
EPASTIM
Mean(S.d.)
60
(23)
159
(106)
451
(120)
729
(228)
877
(184)

% diff.*

-17.6 25.0 18.1 13.8 23.6
a% with reference to EPRRIM values.

TABLE 3. MEANS OF EPRRIM AND EPASTM AT DIFFERENT EPASTM RANGES FOR 90 GLOVE SAMPLES.

EP
(µg/g)
EPRRIM range (µg/g)
50 & less >50 to 100 >100 to 400 >400 to 700 >700
No. of samples 14 16 19 21 20
Corresponding
EPASTM
Range


19 - 260


51 - 267


113 - 835


434 -1017


817 - 1326
EPRRIM
Mean(S.d.)
66
(71)
136
(79)
353
(142)
743
(167)
1065
(162)
EPASTIM
Mean(S.d.)
45
(6)
76
(18)
289
(83)
576
(82)
924
(178)

% diff.*

13.8 44.1 18.1 22.5 13.2
a% difference with reference to EPRRIM values.

Effect of Accelerated Ageing

EPRRIM and EPASTM contents of latex gloves were determined with and without accelerated ageing. Of the 77 lots of commercial gloves tested, 61 lots were powdered gloves, and 16 lots were powder-free of which 4 were siliconized, 3 polymer coated and the rest chlorinated. Results are summarized in Table 4 below.

It is apparent that accelerated ageing resulted in a lowering of 25.2% in the overall means of EPRRIM and 16.5% reduction in the EPASTM measurements. Further analysis of the data showed that the magnitude of changes was relatively greater in the case of the powdered gloves (reduction of 25.4% in EPRRIM, and 16.9% in EPASTM), and less so in the powder-free gloves (10.6% and 4.2% respectively).

The aged EP values were very well related to those of the unaged in both cases, especially in the case of EPRRIM, as illustrated in Figures 2 & 3. The coefficients of correlation are 0.98 for EPRRIM and 0.92 for EPASTM, P < 0.001.

Discussion

Total Extractable Proteins

Although both the RRIM and ASTM tests involved the Lowry colorimetric assay technique, their protocols are not the same. Briefly, the general procedure comprises three parts: (1) Protein extraction, (2) Protein precipitation and (3) Protein quantitation. A number of variations throughout have been adopted by both tests (Appendix). These have consequently resulted in differences not only in the sensitivity of the tests, but also in the extractable protein (EP) values so generated. Sensitivity limits of the methods are, in fact, 20 µg/g and 50 µg/g for the RRIM and the ASTM tests respectively. In agreement with this are also the fact that the EPRRIM values read higher than those of the corresponding EPASTM, for the same latex glove samples tested. Furthermore, it is noteworthy that, as shown by the present findings, low EPRRIM of up to 100 µg/g corresponded to EPASTM of similar range, i.e. up to 107 µg/g (Table 2), while low EPASTM values of 100 µg/g and less were found to be associated with a much wider range of EPRRIM , ie. 267 µg/g and lower (Table 3). Of special interest is the fact that gloves with EPASTM of 50 µg/g (sensitivity limit of the ASTM test) and less, were shown to have EPRRIM varying from 206 µG/g to < 20 µg/g. This could mean that the RRIM test is capable of yielding more sensitive measurements than the ASTM test. On the other hand, it could also be argued that such behavior might be due to differences in response to interference by the two tests. However, there is presently no data to substantiate this. The two sets of EP data are, in fact, very well correlated, coefficient of correlation, r = 0.93, P < 0.001 and n = 90.

Although both tests measure only the total extractable proteins, and not the allergens alone, the EP data so obtained could be of relevance if they relate well to the allergen contents or allergenicity of the test samples. This has, in fact, been shown to be the case for EPRRIM11,12. Gloves with EPRRIM contents varying from > 1000 µg/g to < 20 µg/g were both evaluated for their allergen levels using the IgE ELISA-inhibition test11 , as well as tested by the skin prick test (SPT) on latex hypersensitive subjects from a European population12 . Highly significant relationships were obtained in both cases. More importantly, EPRRIM contents of about 100 µg/g and lower were shown to be associated with not only very low allergen levels, but also with very little allergic response by the sensitive persons. Such findings have rendered the EPRRIM measurements useful as indicators of allergic potential of latex gloves.

There are no similar studies reported on EPASTM, but according to Sussman et.al.13 who recently SPT tested some latex sensitive persons in Canada with gloves of EPASTM contents up to 50 µg/g , more than 50% of the subjects responded positively. In the light of the present observations, it may not be impossible that the gloves tested included some with EPRRIM levels of as high as 200 µg/g, hence contributing to the relatively high positive responses observed. To confirm this, more studies are needed.

TABLE 4. EFFECT OF ACCELERATED AGEING ON EPRRIM AND EPASTM OF LATEX GLOVES

Glove samples EPRRIM(µg/g) EPASTM(µg/g)
Unaged Aged % change Unaged Aged % change
All gloves (n = 77)
Range of EP
Mean EP

5 - 1540
591

3 - 1186
442


- 25.2

14 - 1966
484

8 -1537
404


- 16.5
Powdered gloves (n = 61)
Range of EP
Mean EP

33 - 1540
733

21 - 1186
547


- 25.4

48 - 1966
598

25 - 1537
497


-16.9
Powder-free gloves (n = 16)
Range of EP
Mean of EP

5 - 140
47

3 - 121
42


- 10.6

14 - 159
48

8 - 130
46


- 4.2

Effect of Accelerated Ageing

The Food and Drug Administration (FDA) of the USA, in addressing the latex allergy issue, allowed a "Low Protein Labelling Claim" for latex medical gloves10 in May 1995. As one of the requirements relating to the claim, accelerated ageing is necessary prior to analysis of the test sample for total extractable protein content. This is in view of the fact that the EP content might increase during storage.

The present study has in fact shown that decreases in the extractable protein contents of latex gloves were detected after the ageing process. The drop in EP could likely be due to denaturation of some extractable proteins when the dry gloves were subjected to prolonged heating, leading to certain amount of insolubilization of these proteins. Such decreases were more apparent in the case of the powdered gloves than the powder-free gloves, suggesting that some loss of powder during the ageing process might have partly contributed to this difference.

Very occasional increases in EP have also been observed. The presence of water in the not so dry test sample could be the cause. In such cases, heating during the ageing process could have facilitated further migration of the soluble proteins towards to surface of the film along side with the water, resulting in higher EP contents.

Conclusion

It may be concluded that both the RRIM and the ASTM tests are suitable for measuring total extractable proteins of latex products. Protein values by the former test are generally higher than those of the latter test, and the two sets of data are significantly correlated. However, it should be pointed out that the EP values in both cases are not absolute, being referred to arbitrary protein standards. As such, their comparison could only be deemed more precise if done in reference to a standardized method, which is currently not yet available . But in view of the good correlations between EPRRIM and allergenicity / allergen content, the EPRRIM may be considered to be a useful indicator for allergic potential of latex gloves, particularly for the manufacture of low protein gloves. On the other hand, more data is required in the case of EPASTM.

Accelerated ageing of latex gloves generally has no adverse effect of their extractable protein contents.


Figure 1. Correlation between EP RRIM and correlation EPASTM values of latex gloves.
Coefficient of correlation, r= 0.93, P<0.001, n=90


Figure 2. Relationship between aged and unaged EP RRIM values for 77 lots of latex gloves.
Correlation coefficient, r= 0.98, P<0.001


Figure 3. Relationship between aged and unaged EPASTM values for 77 lots of latex gloves.
Coefficient of correlation, r= 0.92, P<0.001

Acknowledgement

The author wishes to thank the Director of RRIM for permission to present this paper. She also likes to acknowledge the very capable assistance of Miss R. Vijalakshmi, Mr. Ng Chong Seng and En. Abd. Aziz Awang. Contribution by Miss Mok Kok Lang during the initial stage of the study is much appreciated.

1. Proceedings of "International Latex Conference: Sensitivity to latex medical devices", (1992), November 5 - 7, Baltimore, Maryland, USA.

2. Turjanmaa K., Lauren K., Mäkinen- Kiljunen S. and Reunala T. (1988) Rubber contact urticaria: Allergenic properties of 19 brands of latex gloves. Contact Derm., 19, 362.

3. Beezhold D.H. (1993) Measurement of latex protein by chemical and immunological methods. Proceedings of "Latex Protein Allergy: the present position", Amsterdam, 25.

4. Yunginger J.W., Jones R.T., Faraway A.F., Kelso J.M., Warner M.A. and Hunt L.W. (1994) Extractable latex allergens and proteins in disposable medical gloves and other rubber products. J. Allergy Clin. Immunol., 93, 836.

5. Hamilton R.G. and Adkinson N.F. (1994) Serologic assays for antigens and antibodies. Immunol. Allergy Clin. N. Amer., 14, 351.

6. Faridah Yusof and Yeang H.Y. (1992) Quantitation of proteins from natural rubber latex gloves. J. nat. Rubb. Res., 7(3), 206.

7. Test method for the analysis of extractable proteins in natural rubber products. (1996) Malaysian Standard Test, MS 1392 : 96P.

8. Standard test method for analysis of protein in natural rubber and its products (1995) ASTM Designation 5712 - 95.

9. Palosuo T., Mäkinen-Kiljunen S., Alenius H., Reunala T., Esah Yip and Turjanmaa K. (1997) Measurement of natural rubber latex allergen levels in medical gloves by allergen-specific IgE ELISA-inhibition, RAST-inhibition and skin prick testing. Submitted for publication.

10. Interim guidance on protein content labelling claim for latex medical gloves (1995) Food and Drug Administration (FDA), Centre for Devices and Radiological Health (CDRH) and Office of Device Evaluation (ODE) Document.

11. Esah Yip, Palosuo T., Alenius H. and Turjanmaa K. (1997) Correlation between total extractable proteins and allergen levels of natural rubber latex gloves. To be published.

12. Esah Yip, Turjanmaa K., Ng K.P. and Mok K.L. (1995) Proceedings of International Conference on "Latex Protein Allergy: the latest position", Paris, 33.

13. Sussman G.L. and Beezhold D.H. (1997) Safe use of natural rubber latex. Allergy and Asthma Proc. 17(2), 101.

 

APPENDIX 1: RRIM AND ASTM PROTEIN TESTS: MAJOR DIFFERENCES

Mail to:Dr Lai Pin Fah