Calibration Services
Malaysian Rubber Board has an extensive range of calibration equipment for rubber testing. Field of calibration that Malaysian Rubber Board is capable to carry out are as follows:

Instrument Calibrated/ Measurement Parameter
Range
Best measurement capabilities expressed as an uncertainty (±)
Gas measurement
(Site Calibration)

Ozone concentration



0 to 200 pphm



3% of measured value
Mass measurement
(Site-Calibration)

Analytical balance

Weighing device



0 to 200 g

200 g to 4 000 g
4 000 g to 6 000 g
6 000 g to 34 000 g



0.4 mg

3 mg
30 mg
450 mg
Force measurement
(Site calibration)

Tension

Compression



1 N to 500 N

0 to 5 kN
5 kN to 10 kN
10 kN to 100 kN
100 kN to 300 kN
300 kN to 5 000 kN



0.01 % of measured load

0.14 % of measured load
0.28 % of measured load
0.30 % of measured load
0.30 % of measured load
0.74 % of measured load

Instrument Calibrated/ Measurement Parameter
Range
Best measurement capabilities expressed as an uncertainty (±)
Hardness measurement
IRHD Normal (Dead load)
IRHD (Spring load)
JIS A "
Shore A "


0 to 22 N
0 to 22 N
0 to 22 N
0 to 22 N


0.20 % of measured load
0.44 % of measured load
0.81 % of measured load
0.70 % of measured load
Profile Projector
Indentor geometry
- Diameter of indentor
- Indentor extension

Die cutter measurement
- Width
- Length of parallel portion



0 - 90 mm


0 - 90 mm



0.004 mm


0.004 mm
Temperature Heat Enclosure

Type K Thermocouple with Recorder
40°C - 200°C

40°C - 200°C
0.6°C

0.4°C
Plastimeter
Load
Displacement Test
Timing

100N
0.2mm - 1.0 mm
15.0 sec

0.1 N
0.06 mm
0.08 sec
Mooney Viscometer

Torque

Platen Temperature
Rotor Speed

0 to 100 N
(0 to 100 Mooney Unit)

100°C
2 rpm

0.4 N


0.2°C
0.1 rpm
Rheometer (Monsanto R100)

Torque

Timing


Platen Temperature


42 Ib.in

0 min to 40 min
(0 sec to 2400 sec)

160°C
177°C


0.3 Ib.in

0.21sec


0.2°C
0.2°C

Contact


Mr Mohd Nizam Mansor

Technical Manager
Tel :: 03- 6145 9566
Fax : 03-6156 3304
Email : mohdnizam@lgm.gov.my

Calibration Laboratory

Tel : : 03- 6145 9566
Fax : 03-6156 3304
Email : calibration@lgm.gov.my